Polarimeter
A polarimeter is a scientific instrument used to measure optical rotation. This is the angle of rotation caused by passing linearly polarized light through an optically active substance.
Some chemical substances are optically active. When linearly polarized light, which travels in one direction, passes through these substances, it turns either to the left (counter-clockwise) or to the right (clockwise). The amount the light turns is called the angle of rotation.
The direction and amount of this turning reveal information about the sample's chiral properties. For example, they can show the relative amounts of enantiomers present in the sample.
History
See also: Optical rotation § History, and History of crystallography before X-rays § Polarimetry
In 1808, Étienne-Louis Malus (1775–1812) discovered that light can be polarized when it reflects off a surface.
Measuring principle
Polarimetry can be used to measure the ratio, purity, and concentration of two enantiomers. Enantiomers can turn the plane of linear polarized light, which is why they are called optically active. This property is known as optical rotation. Light sources such as a light bulb, a tungsten halogen lamp, or the sun give off light whose electric field points in all directions. Linearly polarized light is different because its waves move in parallel planes. When light passes through a polarizer, only the part that lines up with the polarizer's plane gets through. Optically active substances turn this plane, and the enantiomer is called dextro-rotatory (turning right) or levo-rotatory (turning left), depending on the direction of the turn.
When different enantiomers are together in one solution, their optical activity adds up. Racemates are optically inactive because their left and right turns cancel each other out. The amount of rotation depends on how concentrated the optically active substance is, so polarimeters can measure concentration for pure enantiomer samples. If the concentration is already known, a polarimeter can also help find the specific rotation of a new substance. Optical rotation depends on temperature, concentration, wavelength, path length, and the substance being measured.
Construction
The polarimeter is made up of two Nicol prisms. One is the polarizer, which stays fixed in place. The other is the analyzer, which can be turned. The prisms can be thought of as two slits, S1 and S2, and light waves can be compared to waves moving along a string. The polarizer lets through only the light waves that move in a single plane, so the light becomes plane polarized. When the analyzer is in a similar position, it lets these light waves pass through. When it is turned through a right angle, no waves can get through, and the field looks dark.
If a glass tube holding an optically active solution is placed between the polarizer and the analyzer, the light turns through a certain angle. The analyzer must then be turned through the same angle.
Operation
Polarimeters measure the rotation by passing monochromatic light through the first of two polarizing plates. This first plate is called the polarizer, and it turns the light into a polarized beam. The beam is then rotated as it passes through the sample. After the sample, a second plate called the analyzer is turned, either by hand or automatically, to measure the angle. When the analyzer is turned so that all the light or no light gets through, the angle of rotation can be found. In the first case it equals the angle θ that the analyzer was turned, and in the second case it equals 90-θ.
Types of polarimeter
There are several kinds of polarimeter. Laurent's half-shade polarimeter uses a half-shade device, which divides the field of light into two halves, generally of unequal brightness, so the angle of rotation can be measured more accurately. The biquartz polarimeter uses a biquartz plate made of two semicircular plates of quartz, one right-handed and one left-handed.
The earliest polarimeters, which date back to the 1830s, needed the user to turn one part by hand while looking through another, and to judge the alignment by eye. Most manual polarimeters made today still work this way, but improvements such as a half-wave plate and a long-life yellow LED have made them more accurate and less costly. Semi-automatic polarimeters use a digital display and electronic controls. Fully automatic polarimeters are now widely used. The user presses a button and waits for a digital reading, and they can give continuous measurements for high-performance liquid chromatography. Some modern polarimeters also use a Faraday modulator to find the point of maximum darkness more precisely.
Sources of error
The angle of rotation of an optically active substance can be affected by several things. These include how concentrated the sample is, the wavelength (color) of the light passing through it, and the temperature of the sample. The length of the sample cell also matters, and most automatic polarimeters need the user to enter it to give more accurate results. Air bubbles and uneven temperature or concentration inside the sample can also cause errors.
Generally, the angle of rotation gets smaller as the wavelength gets larger, and it gets larger as the temperature gets higher. Most modern polarimeters have ways to correct or control these errors.
Calibration
Long ago, a sucrose solution of a known strength was used to calibrate polarimeters. The International Commission for Uniform Methods of Sugar Analysis (ICUMSA) helped unify methods for the sugar industry and set standards for polarimeters used there. However, sugar solutions can become contaminated or evaporate, and optical rotation is very sensitive to temperature. A more stable standard was found in crystal quartz, cut so that it matches the optical rotation of a normal sugar solution.
Quartz (silicon dioxide, SiO2) is a common mineral. Today, quartz plates of different thicknesses serve as standards to calibrate polarimeters and saccharimeters. A calibration begins with a preliminary test, and then the optical rotation is measured with a precision polarimeter. The measurement uncertainty of the polarimeter is 0.001°.
Applications
Many optically active chemicals, such as tartaric acid, are stereoisomers. A polarimeter can show which isomer is in a sample: if it turns polarized light to the left, the sample is a levo-isomer, and if it turns light to the right, it is a dextro-isomer. It can also measure the ratio of enantiomers in solutions. The amount of rotation depends on how much of the optically active substance is dissolved, so the instrument can measure concentration for samples made of one enantiomer. Many chemicals have a specific rotation, a unique property like intensive property, refractive index, or specific gravity, which helps identify them. When the specific rotation is known, the concentration or purity of a solution can be calculated, and most automatic polarimeters do this calculation for the user. In the food, beverage, and pharmaceutical industries, these checks help confirm quality for products and ingredients such as steroids, diuretics, antibiotics, narcotics, vitamins, analgesics, amino acids, essential oils, polymers, starches, and sugars.
Polarimeters are also used in the sugar industry to check the quality of juice from sugar cane and of refined sucrose. Sugar refineries often use a modified polarimeter with a flow cell, used together with a refractometer, called a saccharimeter. These instruments use the International Sugar Scale, defined by the International Commission for Uniform Methods of Sugar Analysis (ICUMSA).